White spots, chalky films and stains on freshly sterilised instruments are often caused by water quality rather than the instruments or the autoclave. Fixing the water source usually fixes the problem.

Why water quality matters
Tap water contains dissolved minerals such as calcium and magnesium, and sometimes chlorides. When water evaporates from instruments during drying, these minerals are left behind as spots. Chlorides can also cause pitting corrosion over time.
Recommended water
Most autoclave manufacturers specify distilled, deionised or reverse-osmosis (RO) water. Check your autoclave manual for the required water quality.
Where water matters
- Autoclave reservoir — always use the specified water.
- Final rinse after cleaning — rinsing with treated water prevents mineral spots.
- Ultrasonic solution — using treated water helps the detergent work correctly.
Other causes of spotting
- Detergent residues not rinsed off.
- Instruments not fully dry before sterilisation.
- Autoclave not cleaned or drained as recommended.
- Overloaded chamber preventing proper drying.
Maintenance
- Drain and clean the reservoir as the manufacturer recommends.
- Clean the chamber and door seal regularly.
- Replace filters on schedule.
- Keep a maintenance log.

Removing existing spots
Mineral spots can usually be removed with an instrument cleaner designed for stainless steel. Avoid abrasive pads that scratch the surface.
See also: preventing rust and staining.